SOP for Mettler Analytical Balance H51AR
-
Set all control knobs to zero.
-
Fully release balance (position 1).
-
Line up zero line with pointer using zero-adjustment knob.
-
Arrest balance (position 0).
-
Place tare sheet/container on pan and partially release.
-
Turn 10g knob until arrows appear on filling guide. Turn knob back one step. Repeat with 1g knob.
-
Fully release balance and allow optical scale to stop.
-
Return scale to zero by turning optical tare knob clockwise.
-
Set exact zero with zero-adjustment knob.
-
Arrest balance.
-
Place sample to be weighed on pan.
-
Partially release balance.
-
Repeat step #6.
-
Arrest balance and pause.
-
Repeat step #7.
-
Turn digital counter knob counterclockwise until next lower scale division is in alignment with pointer.
-
Read your result.
-
Arrest balance.