

SOP Owl Horizontal Agarose Gel Electrophoresis Apparatus
-
Prepare gel box for casting:
Set tray in chamber with gasketed ends pressed against the walls of the buffer chamber and insert the comb selected.
-
Cast gel:
Pour cooled gel solution (about 60oC) into tray.
-
Remove tray and rotate:
Remove tray carefully from chamber and turn 90o so that the comb(s) is/are on the same side as the negative electrode (black).
-
Replace tray:
Replace the tray into the chamber, exposing the open ends of the gel to the buffer chamber.
-
Add buffer:
Pour enough running buffer to completely cover the gel; careful rocking of the tray will eliminate any air bubbles under the tray.
-
Remove comb(s):
Carefully remove comb(s) by lifting up slowly. Do not tear or damage wells.
-
Load the samples:
Follow specifications for well bolumes below; in most cases we will be using a 10 tooth comb.
-
Replace lid:
Carefully slide lid onto the unit: the lid electrodes should connect to the chamber electrodes.
-
Connect to power supply and turn on.
Follow protocol for the power supply being used.
-
Cleaning the system:
Use tepid water and soft cloth to clean comb, tray, chamber and lid.
DO NOT use soap or brush on any part of the gel box.
Rinse in deionized water.
|
# Teeth in Comb |
Well Volume |
|
8 |
65ul |
|
10 |
50ul |
|
12 |
40ul |
|
14 |
30ul |
« Biotechnology Laboratory Equipment