SOP for Shimadzu UV-Visible Spectrophotometer
-
Make sure panel is closed and cuvette holders are empty before starting.
-
Turn spectrophotometer "ON" by flipping the yellow switch on the side.
-
The machine will automatically initialize and make a baseline correction.
-
Select "8" CONDITION SET, then press ENTER.
-
Select "5" for LAMP SELECT to turn off UV light bulb, then press ENTER.
-
Press the FILE key.
-
Select "3" for CELLS (E.coli), then press ENTER.
"7" for PICHIA P (yeast)
"8" for CHO cells (animal)
-
It will then ask you for "Parameter Change Y/N", Select NO, and press ENTER.
-
Fill 2 cuvettes each with 2.2 ml of blank solution. When holding the cuvette, please hold it from the line up or it will tamper the measurements. Wipe cuvettes with Kim wipes.
-
Open panel door.
-
Place blank cuvettes in cuvette holders.
-
Press AUTO ZERO key.
-
Remove blank cuvette from front holder and relace with sample to be tested. Close panel door.
-
Press START to take a reading.
-
Remove sample from front holder. Leave machine on for future readings. Only turn off when you are certain that you are done with it.
-
To test additional samples: Place cuvette in front holder and press START for reading. Do not replace cuvette blank in rear holder.
-
Press COPY for a hard copy/printout.
-
Press RETURN to bring you back to Step 8.* (This will erase your old data.)
-
Press FILE to return to the original screen.
-
Remove cuvettes remaining in holders.
-
Flip power switch located on the side to turn the machine off.