CO2 Regulator for REVCO Incubator
See Also: Equipment List
- Start with all valves in the off position.
- Tank valve turned down clockwise completely.
- Regulator valve turned out counterclockwise until it feel free.
- Line valve turned down clockwise completely.
- Slowly open tank valve one full turn counterclockwise.
- Turn regulator valve clockwise until 15 psi is shown on gauge on left.
- Turn line valve counterclockwise until completely open.